IC-ECG guided PICC tip location in a dextrocardia: A case report

Author:

Qiao Weiru1,Qiao Liping2,Li Dongdong3ORCID,Wang Yuanyong1,Ren Yaru1,Liu Qiao1,Wang Ying1,Li Yufang1,Wang Dongdong1,Lv Feifei1,Wu Rong1,Shi Yuan1,Hou Lihua1,Yan Xiaolong1ORCID

Affiliation:

1. Department of Thoracic Surgery, Tangdu Hospital of Air Force Medical University, Xi’an, China

2. Department of Orthopedics, Xianyang First People’s Hospital, Xianyang, China

3. Department of Cardiology, Tangdu Hospital of Air Force Medical University, Xi’an, China

Abstract

Introduction: Mirror people have difficulty with PICC placement due to inversion of organs. Intracavitary electrocardiography (IC-ECG) guided peripherally inserted central catheter (PICC) tip location technique has been widely applied in clinical practice. Case description: Herein, we admitted a 59-year-old man diagnosed with esophageal cancer (EC). Chest X-ray and computed tomography (CT) revealed that the patient was with a mirror-image dextrocardia and situs inversus totalis: the heart and stomach located in the right side of the body, whereas the liver located in the left side. Echocardiography suggested that the apex of the heart pointed toward the right, while the left and right chambers were inverted. The relationship of the heart chambers, structure, and function were normal: left ventricular ejection fraction was 0.67, left atrial diameter was 31 mm, and heart output was 4.7 L/min. Surface ECG showed typical features of a dextrocardia: P-wave inverted on lead I. Amplitude of the R-wave and S-wave decreased gradually on lead chest from V1 to V5. Compared with the normal ECG image, the waves completely exchanged on lead II and III, so as on aVR and aVL. Methods: Column of saline technique can assist operator estimate the tip position in real-time according to P-wave changes. When the height of P-wave reaches to its highest, it means that the tip of catheter has advanced to the target position of cavo-atrial junction (CAJ). Outcomes: Patient was inserted catheter and no adverse events were reported. Conclusions: Measuring the predicted length of catheter is still an indispensible procedure to help precisely adjusting the tip position in IC-ECG guided PICC. Our work provides both supplement for clinical data to facilitate further research and better understanding of special types of PICC to clinicians.

Funder

Tangdu Innovative Development Project

Publisher

SAGE Publications

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3