Influence of arm position change from adduction to abduction on intracavitary electrocardiogram

Author:

Zhu Shu-shu12ORCID,Zhao Jing3,Zhou Xin-yu2,Gao Wei1ORCID,Pan Ming-hao2,Yu Wei-wei1,Hao Hai-xia1,Zhang Hai-jun4ORCID,Yu De-xin5,Wang Qing-yan6,Huang Ping3,Yin Yu-xia7,Wang Lei8

Affiliation:

1. Department of PICC Clinic, Qilu Hospital, Shandong Univerisity, Jinan, Shandong, China

2. School of Nursing, Shandong University, Jinan, China

3. Department of Gynecology, Qilu Hospital, Shandong Univerisity, Jinan, Shandong, China

4. Department of Vascular & Intervention, Tenth People’s Hospital of Tongji University, Shanghai, China

5. Department of Radiology, Qilu Hospital, Shandong Univerisity, Jinan, Shandong, China

6. Department of Health Care, Qilu Hospital, Shandong Univerisity, Jinan, Shandong, China

7. School of Materials Science and Engineering, University of Science and Technology Beijing, Beijing, China

8. Department of Nursing, Beijing Hospital, National Center of Gerontology, Beijing, China

Abstract

Purpose: The aim of this study is to evaluate the influence of arm movements from adduction to abduction on intracavitary electrocardiogram and the position of a catheter tip. Methods: Overall, 192 peripherally inserted central catheter lines were placed under intracavitary electrocardiogram guidance and 188 of them were enrolled in the study. The catheter was first placed at a time point corresponding to the peak P wave with the arm in adduction. The arm was then abducted to 90° without changing catheter insertion length. During the procedure, basal electrocardiogram, intracavitary electrocardiogram, and radiographs with the arm in adduction and abduction were recorded. Amplitude wave changes and catheter movements were measured on electrocardiogram records and radiographs, respectively. Results: In 188 cases, the P wave displayed typical changes, and 97.8% (184/188) catheters were successfully placed correctly. At the peak P wave, the amplitude of the peak P wave was 8.64 times greater than that of the basal P wave, and the P/R ratio was 0.61. When the arm was abducted to 90°, the amplitude of the P wave dropped to 57% of its peak, P/R decreased from 0.61 to 0.34, and the catheter tip moved cephalad 1.00 and 0.77 vertebral body units in male and female patients, respectively. Conclusion: Peripherally inserted central catheter moves toward the heart when the arm position changes from abduction to adduction. Peripherally inserted central catheter tip placement at the peak P wave with patient’s arm in adduction is accurate and can prevent the catheter from advancing too low. R wave can function as a reference for observing P wave changes during peripherally inserted central catheter placement.

Funder

shandong university

Publisher

SAGE Publications

Subject

Nephrology,Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3