Accuracy of dynamic contour tonometry, Goldmann applanation tonometry, and Tono-Pen XL in edematous corneas

Author:

Kontadakis George A.1ORCID,Pennos Alexandros2,Pentari Iro2,Kymionis George D.23,Pallikaris Ioannis G.2,Ginis Harilaos4

Affiliation:

1. Institute of Vision and Optics (IVO), Medical School, University of Crete, 71003 Heraklion, Crete, Greece

2. Institute of Vision and Optics (IVO), Medical School, University of Crete, Heraklion, Greece

3. Jules Gonin Eye Hospital, University of Lausanne, Lausanne, Switzerland

4. Institute of Vision and Optics (IVO), Medical School, University of Crete, Heraklion, Greece Athens Eye Hospital, Athens, Greece

Abstract

Purpose: To investigate in vitro the accuracy of dynamic contour tonometry, Goldmann applanation tonometry, and Tono-Pen XL in edematous corneas. Methods: Experimental study included 20 freshly enucleated porcine eyes. Epithelium was debrided, and eyes were divided in four groups. Groups were immersed in 35%, 40%, 50%, and 60% glycerin solutions for 3 hours. Subsequently, globes were mounted in a special holder, and their intraocular pressure was hydrostatically adjusted. Intraocular pressure was measured by means of dynamic contour tonometry, Goldmann applanation tonometry, and Tono-Pen XL while adjusting true intraocular pressure to 17, 33, and 50 mm Hg. Ultrasound pachymetry was performed. Results: Mean corneal thickness was 914.5 ± 33.3 μm (730–1015 μm). In true intraocular pressure of 33 mm Hg, Goldmann applanation tonometry and dynamic contour tonometry significantly underestimated true intraocular pressure (mean Goldmann applanation tonometry: 14.7 ± 4.8 mm Hg, p < 0.001, mean dynamic contour tonometry: 21.6 ± 6.8, p < 0.001). Tono-Pen XL also underestimated, but difference was not statistically significant (Tono-Pen XL: 27.9 ± 9.7, p = 0.064). In true intraocular pressure of 50 mm Hg, all three methods significantly underestimated (Goldmann applanation tonometry: 17.6 ± 5.3 mm Hg, p < 0.001, dynamic contour tonometry: 26.8 ± 6.3 mm Hg, p < 0.001, Tono-Pen XL: 35.6 ± 8.4 mm Hg, p < 0.001). The error in measured intraocular pressure for each method (true minus measured intraocular pressure) was significantly correlated to true intraocular pressure ( p < 0.001). The intraocular pressure measurements of each eye taken under true intraocular pressure of 17 and 33 mm Hg with the three methods were correlated to each other. Measurements taken under intraocular pressure of 50 mmHg were not correlated to each other. Corneal thickness was not correlated to intraocular pressure measurement. Conclusion: Goldmann applanation tonometry, dynamic contour tonometry, and Tono-Pen XL underestimate intraocular pressure when measured under edematous conditions. Tono-Pen XL showed better accuracy, especially in lower true intraocular pressure. The measurement error increases when true intraocular pressure increases in all three methods.

Funder

Dr George A. Kontadakis was supported by a grant from the “Hellenic Society of Intraocular Implant and Refractive Surgery”

Publisher

SAGE Publications

Cited by 6 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3