Background. The current articulating instruments used in laparoendoscopic single-site surgery do not appear to provide the joint forces required. Thus, we measured the joint forces of first-generation articulating laparoscopic instruments. To compare these forces with those necessary in the surgical context, we evaluated the forces sufficient to produce secure surgical ties in an animal model. Methods. The articulating instruments tested were Laparo-Angle (Cambridge Endoscopic Devices Inc, Framingham, MA), RealHand (Novare Surgical Systems Inc, Cupertino, CA), and Roticulator (Covidien Inc, Mansfield, MA). For each, we measured the angle between the end-effector and the shaft in proportion to the articulating force using a push–pull gauge. Two fixed-position configurations of the instruments were predetermined: the neutral and the fully articulated positions. The forces required to secure surgical ties for the ureter, renal artery, and renal vein were evaluated using kidneys harvested from a female pig. Results. The bending forces required to bend from the neutral position to 30° were 5.6 ± 1.2 and 4.7 ± 1.0 N with the Laparo-Angle and RealHand, respectively. Furthermore, the slippage forces in the fully articulated state were 1.8 ± 0.3, 1.6 ± 0.2, and 1.5 ± 0.2 N in the above order. In contrast, the mean forces to produce surgical ties of the ureter, renal artery, and renal vein were 14.5 ± 2.3, 11.5 ± 0.8, and 10.3 ± 2.3 N, respectively. Conclusion. The joint forces of first-generation articulating instruments for laparoendoscopic single-site surgery are not sufficient to meet the usual operative needs. Improved articulating instruments with greater articulating forces should be developed.
Cited by
16 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献