New Backstrap Vagus Electrode for Continuous Intraoperative Neuromonitoring in Thyroid Surgery

Author:

Lamadé Wolfram1,Ulmer Christoph1,Rieber Fabian1,Friedrich Colin1,Koch Klaus P.2,Thon Klaus-Peter1

Affiliation:

1. Robert-Bosch Hospital, Stuttgart, Germany

2. University of Applied Sciences Trier, Trier, Germany

Abstract

Introduction. Continuous intraoperative neuromonitoring (CIONM) via vagal nerve stimulation allows real-time surveillance of the recurrent laryngeal nerve during thyroid surgery. However, for effective CIONM, subtle changes in recurrent laryngeal nerve conductivity have to be detected. A newly developed stimulation electrode that provides stable nerve stimulation and safe application is presented. Methods. For electrode validation, current distribution was simulated with the finite element method. Mechanical characteristics were assessed through bench testing. Clinical evaluation was initiated with 11 thyroid surgeries. Results. Experimental and clinical results led to the development of a tripolar gold/polyimide electrode mounted onto a backstrap-shaped silicone body. It facilitated rapid electrode implantation and extraction (median implantation time 4 ± 19 seconds). Peak extraction force was 570 mN. Median supramaximal stimulation currents were 2.00 ± 0.95 mA and resulted in reliable electromyogram responses (median 3.1 ± 3.0 mV). No intraoperative electrode dislocations occurred, and no postoperative nerve palsy was observed. Conclusion. The new backstrap vagal stimulation electrode meets the requirements for reliable CIONM.

Publisher

SAGE Publications

Subject

Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3