Affiliation:
1. University of Rochester Rochester, NY 14627
Abstract
The statistics of speckle patterns in ultrasound images have potential for tissue characterization. In “fully developed speckle” from many random scatterers, the amplitude is widely recognized as possessing a Rayleigh distribution. This study examines how scattering populations and signal processing can produce non-Rayleigh distributions. The first order speckle statistics are shown to depend on random scatterer density and the amplitude and spacing of added periodic scatterers. Envelope detection, amplifier compression, and signal bandwidth are also shown to cause distinct changes in the Signal distribution.
Subject
Radiology Nuclear Medicine and imaging,Radiological and Ultrasound Technology
Cited by
101 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献