The purpose of this study was to quantitate the influence of bright intramural echoes on estimates of myocardial attenuation from analyses of backscattered ultrasound. To achieve this, M-mode image-based measurements of the inherent anisotropic properties of myocardial attenuation were performed on rotating myocardial specimens. The approach was to use a commercially-available ultrasonic imaging system to acquire M-mode images of 24 excised cylindrical specimens from six formalin-fixed lamb hearts for data analysis using a video signal analysis technique. As a control, through-transmission rf-based measurements were performed concurrently using a pair of focused, single-element ultrasonic transducers. We devised an objective approach to compensate M-mode results for the presence of bright intramural echoes that makes use of the rotational symmetry of the measurements. A comparison of the uncompensated and compensated estimates of attenuation shows that the effect of bright intramural echoes under the conditions of this study increases the average error in M-mode results by approximately 240% compared with that observed when such effects are minimized by compensation. For both uncompensated and compensated M-mode results, increased temporal averaging shows only a modest reduction in average error. These data suggest that for measurements of attenuation from backscattered ultrasound using M-mode images, the effects of bright intramural echoes can be a significant source of error despite increased temporal averaging and therefore may require compensation.
Subject
Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献