Combined ARFI and Shear Wave Imaging of Prostate Cancer: Optimizing Beam Sequences and Parameter Reconstruction Approaches

Author:

Chan Derek Y.1ORCID,Morris Daniel Cody1,Polascik Thomas J.2,Palmeri Mark L.1,Nightingale Kathryn R.1ORCID

Affiliation:

1. Department of Biomedical Engineering, Duke University, Durham, NC, USA

2. Department of Urology, Duke University Medical Center, Durham, NC, USA

Abstract

This study demonstrates the implementation of a shear wave reconstruction algorithm that enables concurrent acoustic radiation force impulse (ARFI) imaging and shear wave elasticity imaging (SWEI) of prostate cancer and zonal anatomy. The combined ARFI/SWEI sequence uses closely spaced push beams across the lateral field of view and simultaneously tracks both on-axis (within the region of excitation) and off-axis (laterally offset from the excitation) after each push beam. Using a large number of push beams across the lateral field of view enables the collection of higher signal-to-noise ratio (SNR) shear wave data to reconstruct the SWEI volume than is typically acquired. The shear wave arrival times were determined with cross-correlation of shear wave velocity signals in two dimensions after 3-D directional filtering to remove reflection artifacts. To combine data from serially interrogated lateral push locations, arrival times from different pushes were aligned by estimating the shear wave propagation time between push locations. Shear wave data acquired in an elasticity lesion phantom and reconstructed using this algorithm demonstrate benefits to contrast-to-noise ratio (CNR) with increased push beam density and 3-D directional filtering. Increasing the push beam spacing from 0.3 to 11.6 mm (typical for commercial SWEI systems) resulted in a 53% decrease in CNR. In human in vivo data, this imaging approach enabled high CNR (1.61–1.86) imaging of histologically-confirmed prostate cancer. The in vivo images had improved spatial resolution and CNR and fewer reflection artifacts as a result of the high push beam density, the high shear wave SNR, the use of multidimensional directional filtering, and the combination of shear wave data from different push beams.

Funder

National Institutes of Health

U.S. Department of Defense

Publisher

SAGE Publications

Subject

Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Ultrasound Viscoelastography by Acoustic Radiation Force: A State-of-the-Art Review;IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control;2024-05

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3