Abstract
Objective In the present study, 50 EVAS procedures were evaluated in regard to primary (survival and technical success) and secondary (device-related complications) events of interest. Methods The single center study was conducted from July 2013 to August 2014 with prospective collection of the clinical data. The clinical results were controlled by CT angiography and contrast-enhanced ultrasound . Results The technical success was 98% and the 30-day mortality 4%. One (2%) patient died from multisystem organ failure and another patient from an intracranial bleeding, respectively. One patient (2%) suffered from a device-related aneurysm rupture. During early follow-up, one (2%) patient developed an endoleak type II, while three (6%) patients suffered from a partial endograft limb thrombosis. Overall, a secondary intervention was necessary in six (12%) patients. Conclusions With the Nellix EVAS system, a high primary technical success of 98% was achieved; one (2%) patient developed an endoleak type II which did not require secondary intervention. Those promising results are contrasted by a substantial rate of endograft limb thromboses (8%) and one (2%) intraoperative aneurysm rupture. Further studies are needed to assess the durability of the Nellix stentgraft and the occurrence of device-related complications.