Affiliation:
1. Vascular and Endovascular Surgery Unit, University of Rome Tor Vergata, Roma, Italy
2. Vascular and Endovascular Surgery Unit, Sandro Pertini Hospital, Rome, Italy
Abstract
Purpose To describe the off-label use of tapered iliac limbs for the treatment of isolated iliac aneurysms with proximal landing zone significantly larger than distal landing zone. Technique Inversion of a Gore Excluder tapered leg (W. L. Gore & Associates Inc, Flagstaff, Arizona) with a modified upside-down technique is described. The endoprosthesis, with the olive at the tip of the releasing system previously cut, is inserted in a tip-to-tip fashion into a 15 Fr introducer sheath. The graft is released inside the introducer. An 18 Fr introducer sheath is advanced up to the proximal sealing zone. Following the removal of the 18 Fr dilator, the 15 Fr introducer with the pre-released graft is inserted co-axially into the 18 Fr introducer. A pre-cut 15 Fr dilator is brought up to the endograft and used as a pusher. A pull-back maneuver of the co-axial system, countertractioning with the dilator maintained in position, allows the delivery of the endograft. Conclusion This technique might offer a feasible option in case of endovascular exclusion of isolated iliac artery aneurysms with significant landing zone diameter mismatch. Extracorporeal inversion is time-saving and could be safer in terms of graft damage and infection.