AbsorbaSeal™ 5.6.7F vascular closure device: A good laboratory practice chronic study evaluating safety and efficacy in a healthy porcine model

Author:

Ayuso Sullivan A1ORCID,Shipp R Caroline2,Aladegbami Bola G1,Farquharson Delton3,Lawson Denny4,Bassett Randy4

Affiliation:

1. Division of Gastrointestinal and Minimally Invasive Surgery, Department of Surgery, Carolinas Medical Center, Charlotte, NC, USA

2. The University of Tennessee at Chattanooga College of Arts and Sciences, Chattanooga, TN, USA

3. General and Vascular Surgery, Princess Margaret Hospital, Nassau, Bahamas

4. CyndRx, LLC, Brentwood, TN, USA

Abstract

Objectives Vascular closure devices (VCDs) are widely used for arteriotomy closure after percutaneous catheter–based procedures. In comparison to manual compression, VCDs have been associated with shorter time to hemostasis, shorter time to ambulation, and also decreased length of stay. Complexity of deployment, lack of immediate hemostasis, and residual deformity of the arterial wall remain as limitations of current VCDs. The aim of this study was to investigate the AbsorbaSeal™ 5.6.7F vascular closure device, a novel, completely bioabsorbable, intuitive, and easy to use VCD which uses a compressive, “sandwich”-type design comprising a low profile intravascular distal seal and gasket and an extravascular floating foot and proximal seal, in an open infrarenal aortic swine model. Methods Eight fully heparinized swine at a good laboratory practices facility underwent AbsorbaSeal™ 5.6.7F VCD closure of three 6F arteriotomies each in the proximal, mid, and distal infrarenal aorta. Two swine underwent harvest at each of four time cohorts: 30, 60, 90, and 120 days. Acute and chronic procedural safety and efficacy, as well as target site vascular remodeling over time, were the primary outcomes evaluated. Secondary outcome measures included local and systemic inflammatory responses, end-organ tissue analysis, and device-related complications through the follow-up periods. Histopathological evaluation was performed by a blinded pathologist. Standard statistical methods were used. Results In deployment of 24 AbsorbaSeal™ 5.6.7F VCDs, there were no device-related complications or mortalities. All deployments resulted in rapid arteriotomy seal (100% deployment success), with a mean time to hemostasis (cessation of arterial flow) of 21.5 s (median: 6.5 s) across a mean activated clotting time (ACT) of 356 s. Twenty of the 24 implant sites (83%) attained complete hemostasis within 20 s. Immediate post-implant and pre-termination angiographies at all time points were performed of all swine which demonstrated normal aortic appearance and tissue structure and normal downstream vascular beds. At 30 days, each implant’s intravascular distal seal and gasket were removed from the circulation and completely covered with a smooth neointimal layer. Minimal inflammation and no intimal or luminal thrombus were observed at any site at every time point. Conclusions AbsorbaSeal™ 5.6.7F is a safe, effective, and secure VCD that demonstrates rapid hemostasis in a fully heparinized open aortic porcine model. Removal from circulation and complete coverage of the intravascular distal seal and gasket with neointima occurred within 30 days post-implant. Natural transmural vessel healing from the arteriotomy itself with minimal inflammation was noted for each implant at every time point.

Funder

CyndRx

Publisher

SAGE Publications

Subject

Cardiology and Cardiovascular Medicine,Radiology, Nuclear Medicine and imaging,General Medicine,Surgery