Point-of-care coagulation testing for reducing in-hospital delay in thrombolysis

Author:

Han Jung Hee1,Jang Seongsoo2,Choi Mi-Ok2,Yoon Mi-Jeong1,Lim Seung-Bok1,Kook Jeong-Ran1,Kang Dong-Wha3,Kwon Sun U3,Kim Jong S3,Jeon Sang-Beom3

Affiliation:

1. Department of Nursing, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Republic of Korea

2. Department of Laboratory Medicine, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Republic of Korea

3. Department of Neurology, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Republic of Korea

Abstract

Background: The confirmation of prothrombin time international normalized ratio by a central laboratory often delays intravenous thrombolysis in patients with acute ischemic stroke. Objectives: We investigated the feasibility, reliability, and usefulness of point-of-care determination of prothrombin time international normalized ratio for stroke thrombolysis. Methods: Among 312 patients with ischemic stroke, 202 who arrived at the emergency room within 4.5 h of stroke onset were enrolled in the study. Patients with lost orders for point-of-care testing for the prothrombin time international normalized ratio or central laboratory testing for the prothrombin time international normalized ratio (n = 47) were excluded. We compared international normalized ratio values and the time interval from arrival to the report of test results (door-to-international normalized ratio time) between point-of-care testing for the prothrombin time international normalized ratio and central laboratory testing for the prothrombin time international normalized ratio. In patients who underwent thrombolysis, we compared the time interval from arrival to thrombolysis (door-to-needle time) between the current study population and historic cohort at our center. Results: In the 155 patients included in the study, the median door-to-international normalized ratio time was 9.0 min (interquartile range, 5.0–12.0 min) for point-of-care testing for the prothrombin time international normalized ratio and 46.0 min (interquartile range, 38.0–55.0 min) for central laboratory testing for the prothrombin time international normalized ratio (p < 0.001). The intraclass correlation coefficient between point-of-care testing for the prothrombin time international normalized ratio and central laboratory testing for the prothrombin time international normalized ratio was 0.975 (95% confidence interval: 0.966–0.982). Forty-nine of the 155 patients underwent intravenous thrombolysis. The door-to-needle time was significantly decreased after implementation of point-of-care testing for the prothrombin time international normalized ratio (median, 23.0 min; interquartile range, 16.0–29.8 vs median, 46.0 min; interquartile range, 33.5–50.5 min). Conclusion: Utilization of point-of-care testing for the prothrombin time international normalized ratio was feasible in the management of patients with acute ischemic stroke. Point-of-care testing for the prothrombin time international normalized ratio was quick and reliable and had a pivotal role in expediting thrombolysis.

Publisher

SAGE Publications

Subject

Emergency Medicine

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3