Decentralized Clinical Chemistry Testing: Quality of Results Obtained by Residents and Interns in an Acute Care Setting

Author:

Nanji Amin A.1,Poon Raymond2,Hinberg Irwin2

Affiliation:

1. Departments of Pathology and Laboratory Medicine, University of Ottawa and Ottawa General Hospital, Department of Laboratory Medicine, Ottawa General Hospital, 501 Smyth Rd, Ottawa, Ontario, K1H 8L6 Canada.

2. Bureau of Radiation and Medical Devices, Health Protection Branch, Ottawa, Ontario, Canada

Abstract

We evaluated the quality of results obtained for near- patient testing by residents and interns (MDs) using desk-top analyzers. The instruments evaluated were the Reflotron analyzer, the Seralyzer, the Vision analyzer, and the Kodak DT60 analyzer. Low- and high-concen trations of control materials were used for precision studies. Results obtained by MDs were compared with those obtained by a trained technologist for comparison studies. The range of precision obtained by MDs was 4.6 to 6.6% for the Reflotron analyzer, 3.6 to 20.7% for the Seralyzer, 1.7 to 5.3% for the Vision analyzer, and 2.7 to 13% for the Kodak DT60 analyzer. The range of correla tion coefficients for each of the analyzers was 0.980 to 0.994 for the Reflotron analyzer, 0.763 to 0.933 for the Seralyzer, 0.989 to 0.999 for the Vision analyzer, and 0.987 to 0.999 for the Kodak DT60. The percentage of results for each of the analyzers when the results ob tained by MDs were more than 10% different from the trained technologist's results were 17.3% for the Reflo tron, 44% for the Seralyzer, 0% for the Vision, and 17% for the Kodak DT60. The discrepancies between MD and the trained technologist's results are related to the complexity of the instruments used.

Publisher

SAGE Publications

Subject

Critical Care and Intensive Care Medicine

Cited by 7 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Point of Care Testing in Critical Care;Critical Care Nursing Clinics of North America;1999-03

2. New Whole Blood Analyzers and Their Impact on Cardiac and Critical Care;Critical Reviews in Clinical Laboratory Sciences;1993-01

3. Seralyzer;Dry Chemistry - Analysis with Carrier-Bound Reagents;1993

4. Reflotron;Dry Chemistry - Analysis with Carrier-Bound Reagents;1993

5. Ektachem;Dry Chemistry - Analysis with Carrier-Bound Reagents;1993