Affiliation:
1. University of Illinois College of Medicine at Peoria,
2. Children’s Hospital of Illinois at OSF St. Francis Medical Center, Peoria, Illinois.
3. University of Illinois College of Medicine at Peoria
Abstract
The objective of this prospective, observational study with consecutive sampling was to assess the reliability, bias, and precision of Nellcor N-395 (N) and Masimo SET Radical (M) pulse oximeters in children with cyanotic congenital heart disease and children with congenital heart disease recovering from cardiopulmonary bypass-assisted surgery admitted to a cardiovascular operating suite and pediatric intensive care unit at a tertiary care community hospital. Forty-six children with congenital heart disease were studied in 1 of 2 groups: (1) those recovering from cardiopulmonary bypass with a serum lactic acid > 2 mmol/L, and (2) those with co-oximetry measured saturations (SaO 2) < 90% and no evidence of shock. Measurements of SaO 2 of whole blood were compared to simultaneous pulse oximetry saturations (SpO 2). Data were analyzed to detect significant differences in SpO 2 readout failures between oximeters and average SpO 2 - SaO 2 ± 1 SD for each oximeter. A total of 122 SaO 2 measurements were recorded; the median SaO 2 was 83% (57 - 100%). SpO 2 failures after cardiopulmonary bypass were 41% (25/61) for N versus 10% (6/61) for M ( P < .001). There was a significant difference in bias (ie, average SpO 2 - SaO 2) and precision (± 1 SD) between oximeters (N, 1.1 ± 3.3 vs M, -0.2 ± 4.1; P < .001) in the postcardiopulmonary bypass group but no significant difference in bias and precision between oximeters in the cyanotic congenital heart disease group (N, 2.9 ± 4.6 vs M, 2.8 ± 6.2; P = .848). The Nellcor N-395 pulse oximeter failed more often immediately after cardiopulmonary bypass than did the Masimo SET Radical pulse oximeter. SpO2 measured with both oximeters overestimated SaO2 in the presence of persistent hypoxemia.
Subject
Critical Care and Intensive Care Medicine
Cited by
19 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献