A Clinical Device for Measuring Internal-External Rotational Laxity of the Knee

Author:

Alam Mahbub1234,Bull Anthony M.J.2,Thomas Rhidian deW3,Amis Andrew A.12

Affiliation:

1. Department of Mechanical Engineering, Imperial College London, London, United Kingdom

2. Department of Bioengineering, Imperial College London, London, United Kingdom

3. Department of Musculoskeletal Surgery, Imperial College London, London, United Kingdom

4. Department of Orthopaedic Surgery, Royal National Orthopaedic Hospital, Stanmore, United Kingdom

Abstract

Background:The dial test can be improved by providing reproducible, accurate measurements to improve diagnosis and treatment comparisons.Purpose:Validation of a rotational measurement device (RMD) for measuring knee internal-external rotational laxity.Study Design:Cohort study (diagnosis); Level of evidence, 2.Methods:The RMD consisted of 3 components: a femoral clamp and a tibial splint using paired inclinometers to measure rotations, and a boot to apply tibial internal-external rotation torque. A separate boot inclinometer allowed for foot rotations to be measured independently. The measurements were simultaneously compared with electromagnetic “nest of birds” (NOB) sensors. Sequential paired knee measurements were taken at 4, 6, and 8 N·m of torque at 30° and 90° of flexion in 46 volunteers.Results:The correlation coefficient was 0.92 (95% CI, 0.89 to 0.94) and 0.63 (95% CI, 0.54 to 0.70) between the NOB and RMD and between the NOB and boot inclinometer, respectively. Bland-Altman analysis revealed that the RMD was on average within 2° (95% CI, 1° to −4°) of NOB readings, whereas the boot overestimated by 34° (95% CI, −9° to −58°). Maximum side-to-side differences measured by the NOB, RMD, and boot were 1°, 3°, and 21°, respectively. The mean +2× standard deviation data gave a range of side-to-side differences of less than 5° for the RMD. The intraobserver intraclass correlation was 0.9 (95% CI, 0.78 to 0.97) at both 30° and 90° of flexion, and the 95% CI of the differences between readings taken on 2 occasions, the interobserver repeatability, was 1° or less.Conclusion:The novel clinical RMD for measuring rotational laxity of the knee was portable, easy, and comfortable to use in the clinical setting. The RMD showed significant correlation and accuracy compared with sensors of known high accuracy. Side-to-side differences of less than 5° were found in 95% of normal knees, compared with differences of 13° or more for clinical diagnosis of pathological rotational laxity. Measuring knee rotation at the foot showed poor correlation and accuracy.

Publisher

SAGE Publications

Subject

Physical Therapy, Sports Therapy and Rehabilitation,Orthopedics and Sports Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3