The Biomechanical Stability of Bone Staples in Cortical Fixation of Tendon Grafts for Medial Collateral Ligament Reconstruction Depends on the Implant Design

Author:

Deichsel Adrian1,Raschke Michael J.1,Herbst Elmar1,Peez Christian1,Oeckenpöhler Simon1,Briese Thorben1,Wermers Jens1ORCID,Kittl Christoph1,Glasbrenner Johannes1

Affiliation:

1. Department of Trauma, Hand and Reconstructive Surgery, University Hospital Münster, Münster, Germany

Abstract

Background:The promising biomechanical stability of bone staples (BSs) in cortical fixation of tendon grafts for medial collateral ligament (MCL) reconstruction has been revealed by a previous investigation. However, it is currently unknown if the biomechanical stability of cortical fixation of tendon grafts depends on the BS design.Purpose:To assess the biomechanical stability of cortical fixation of tendon grafts in knee surgery using 4 different BS designs.Study Design:Controlled laboratory study.Methods:Cortical fixation of tendon grafts was performed in a porcine knee model at the tibial insertion area of the MCL using 4 different BS designs (n = 40): 8-mm width without spikes (n = 10), 8-mm width with spikes (n = 10), 14-mm width with spikes (n = 10), and 13 mm–wide 4-prong staples with spikes (n = 10). Specimens were mounted in a materials testing machine, and cyclic loading was applied to the tendon graft (500 cycles at 50 and 100 N, respectively), followed by load-to-failure testing. The Kruskal-Wallis test was performed for statistical analysis ( P < .05), and the post hoc Dunn test was performed for multiple comparisons.Results:In 4 of 10 specimens with graft fixation using BSs without spikes, slippage of the tendon underneath the BS led to failure of the construct during cyclic loading to 100 N. In the other groups, no fixation failure was observed during cyclic loading. Furthermore, graft fixation using BSs without spikes was found to have significantly more elongation during cyclic loading (8.2 ± 1.9 mm) and a lower ultimate failure load (170 ± 120 N) compared with graft fixation using narrow BSs with spikes (3.4 ± 1.2 mm [ P < .0001] and 364 ± 85 N [ P < .05], respectively) and graft fixation using broad BSs with spikes (4.5 ± 1.4 mm [ P < .05] and 429 ± 67 N [ P < .001], respectively). No statistical differences in elongation during cyclic loading or ultimate failure load were found between 4-prong staples with spikes (5.0 ± 1.3 mm and 304 ± 85 N) and narrow or broad staples with spikes.Conclusion:The biomechanical stability of cortical fixation of an MCL graft was comparable between each BS design with spikes (narrow, broad, and 4-prong) in a porcine knee model, whereas BSs without spikes led to failure of the fixation construct during cyclic loading in 4 of 10 specimens and increased elongation and lower ultimate failure loads in the remainder of the group. BSs without spikes may therefore not be recommended for graft fixation.Clinical Relevance:The use of BSs can help to avoid the conflict of converging tunnels in multiligament reconstruction surgery. An implant design with spikes yields significantly higher biomechanical stability than BSs without spikes.

Publisher

SAGE Publications

Subject

Physical Therapy, Sports Therapy and Rehabilitation,Orthopedics and Sports Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3