Cadaveric Biomechanical Study of Partial Glenoid Arthroplasty Versus the Latarjet Procedure for Anterior Glenoid Bone Loss

Author:

Kim Yong Tae1ORCID,Lee Kyung Jae2,Jang Young Hoon3,Yang Sook4,Lee Thay Q.5,McGarry Michelle5ORCID,Kim Sae Hoon6ORCID

Affiliation:

1. Department of Orthopedic Surgery, Hallym University Dongtan Sacred Heart Hospital, Hwaseong, Republic of Korea

2. Department of Orthopedic Surgery, Chung-Ang University Hospital, Seoul, Republic of Korea

3. Department of Orthopedic Surgery, CM Hospital, Seoul, Republic of Korea

4. Research Center, Cusmedi, Suwon, Republic of Korea

5. Orthopaedic Biomechanics Laboratory, Congress Medical Foundation, Pasadena, California, USA

6. Department of Orthopedic Surgery, Seoul National University Hospital, Seoul, Republic of Korea

Abstract

Background: For severe anterior glenoid bone loss due to recurrent shoulder instability, the Latarjet procedure offers a dynamic sling effect in addition to bone augmentation. Yet, it heavily alters the surrounding anatomy, while fixation and graft union issues are also common. Purpose/Hypothesis: The purpose of this study was to compare a novel printed 3-dimensional (3D) partial glenoid arthroplasty (PGA) implant with the classic Latarjet procedure. It was hypothesized that by replicating the original glenoid geometry and preserving soft tissue anatomy, PGA may better reproduce normal joint kinematics. In addition, the locking screw construct may offer stronger fixation. Study Design: Controlled laboratory study. Methods: A total of 14 matched cadaveric shoulders were tested. The PGA implant was 3D printed in titanium based on preoperative computed tomography. The intact, 25% anterior glenoid bone loss, and postoperative states were tested in the scapular and coronal planes. The following parameters were measured: articular surface area and stepoff, rotational range of motion and the humeral head apex position during rotation, and load and linear stiffness at 25% anterior translation and at 2-mm construct displacement. Results: The baseline dimensions of the glenoid articular surface were comparable between the groups. The articular surface area after PGA was significantly larger ( P = .006) with less articular stepoff ( P = .030). PGA better approximated the intact state’s external ( P = .006) and total ( P = .019) rotational range of motion in the scapular plane. The course of the humeral head apex after PGA better followed that of the intact state ( P < .001). Resistance against anterior translation after PGA was not significantly different compared with after the Latarjet procedure. Greater linear stiffness ( P = .031) and loading ( P = .002) at 2-mm construct displacement were demonstrated in the PGA group. Conclusion: In addressing anterior glenoid bone loss, PGA better approximated intact glenohumeral joint kinematics compared with the Latarjet procedure with less articular stepoff in a cadaveric model. PGA was comparable in resisting anterior translation while being significantly stronger against loading at 2-mm construct displacement. Further clinical studies are warranted to validate this novel procedure. Clinical Relevance: A 3D-printed PGA implant may offer an alternative treatment option for severe glenoid bone loss due to shoulder instability, overcoming the previous drawbacks of the Latarjet procedure, including altered kinematics, fixation failure, and hardware issues.

Funder

Korea Medical Device Development Fund

Publisher

SAGE Publications

Subject

Physical Therapy, Sports Therapy and Rehabilitation,Orthopedics and Sports Medicine