Author:
Rorabeck C.H.,Kennedy J.C.
Abstract
Five patients developed paralysis to the sciatic nerve after the application of a pneumatic tourniquet for knee ligament sur gery. Four of these patients made a complete neurologic recov ery within 6 months. A subsequent experimental investigation has demonstrated that some degree of impaired function in the sciatic nerve of a dog occurs with every tourniquet application. On the basis of this experimental work, it is recommended that a pneumatic tourniquet not be left inflated for periods longer than 1 hr and 15 min. In addition, it is suggested that once a tourniquet has been deflated during an operative procedure, that it should not be reinflated because of the excessive length of time required for the sciatic nerve conduction velocity to return to normal. Tourniquets should also be checked for calibration before use at time of an operation.
Subject
Physical Therapy, Sports Therapy and Rehabilitation,Orthopedics and Sports Medicine
Cited by
110 articles.