In Vivo Performance of a Novel, Anatomically Shaped, Total Meniscal Prosthesis Made of Polycarbonate Urethane: A 12-Month Evaluation in Goats

Author:

Vrancken Anne C.T.1,Hannink Gerjon1,Madej Wojciech1,Verdonschot Nico12,van Tienen Tony G.13,Buma Pieter1

Affiliation:

1. Department of Orthopaedics, Radboud University Medical Center, Nijmegen, the Netherlands

2. Laboratory of Biomechanical Engineering, University of Twente, Enschede, the Netherlands

3. Department of Orthopaedic Surgery, Kliniek ViaSana, Mill, the Netherlands

Abstract

Background: Injury or loss of the meniscus generally leads to degenerative osteoarthritic changes in the knee joint. However, the treatment options for symptomatic patients with total meniscectomy are limited. Therefore, we developed a novel, anatomically shaped, total meniscal implant made of polycarbonate urethane. Purpose: To evaluate the in vivo performance of this novel total meniscal implant. The assessment particularly focused on the implant’s response to long-term physiological loading in a goat model and its chondroprotective capacity in comparison to clinically relevant controls. Study Design: Controlled laboratory study. Methods: Surgery was performed to the stifle joint of 26 female Saanen goats, subdivided into 4 groups: implant, allograft, total meniscectomy, and sham surgery. The sham group’s contralateral joints served as nonoperated controls. After 12 months of follow-up, investigators evaluated implant wear, deformation, and the histopathological condition of the synovium and cartilage. Results: Wear of the implant’s articulating surfaces was minimal, which was confirmed by the absence of wear particles in the synovial fluid. Implant deformation was limited. However, one implant failed by complete tearing of the posterior horn extension. No differences in cartilage histopathological condition were observed for the implant, allograft, and meniscectomy groups. However, locally, the cartilage scores for these groups were significantly worse than those of the nonoperated controls. Conclusion: Whereas this study demonstrated that the novel implant is resistant to wear and that deformation after 12 months of physiological loading is acceptable, reinforcement of the implant horns is necessary to prevent horn failure. Although the implant could not protect the cartilage from developing degenerative changes, the progression of damage was similar in the allograft group. Clinical Relevance: This novel polycarbonate urethane implant may have the potential to become an alternative treatment for symptomatic patients with total meniscectomy.

Publisher

SAGE Publications

Subject

Physical Therapy, Sports Therapy and Rehabilitation,Orthopedics and Sports Medicine

Cited by 18 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3