Comparison of Measurements of Implant Stability by Two Different Radio Frequency Analysis Systems: An In Vitro Study

Author:

Koyuncuoglu Cenker Zeki1ORCID,Demir Becen2

Affiliation:

1. Faculty of Dentistry, Department of Periodontology, Istanbul Aydin University, Istanbul, Turkey

2. Private Practice, Istanbul, Turkey

Abstract

Aim: To compare and evaluate the implant stability quotient (ISQ) measurements by two different radio frequency analysis (RFA) machines, Osstell Mentor and PenguinRFA. Materials and Methods: Twenty bone-level implants (3.7 × 10 mm2) were placed on a bovine bone in this in vitro study. The ISQ measurements were performed with the PenguinRFA and Osstell Mentor machines using the multipeg of the Penguin system. The measurements were repeated three times for each direction and an average ISQ value was calculated. The results were averaged (mean ± SD) and the intraclass correlation coefficient (ICC) was calculated to assess the relationship between the measurements. Result: The mean ISQ values for the Osstell and Penguin machines were 77.60 + 2.11 and 78.05 + 2.04, respectively. The ICC was 0.958. The ISQ values obtained from the Osstell and Penguin machines were significantly compatible ( P < .05). Conclusion: Both of the RFA devices provided similar and reliable ISQ measurements.

Publisher

SAGE Publications

Subject

Pharmacology (medical)

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3