Affiliation:
1. Institute of Biostatistics, Leibniz Universität Hannover, Hannover, Germany
Abstract
When a new in vitro assay method is introduced, it should be validated against the best available knowledge or a reference standard assay. For assays resulting in a simple binary outcome, the data can be displayed as a 2 × 2 table. Based on the estimated sensitivity and specificity, and the assumed prevalence of true positives in the population of interest, the positive and negative predictive values of the new assay can be calculated. We briefly discuss the experimental design of validation experiments and previously published methods for computing confidence intervals for predictive values. The application of the methods is illustrated for two toxicological examples, by using tools available in the free software, namely, R: confidence intervals for predictive values are computed for a validation study of an in vitro test battery, and sample size calculation is illustrated for an acute toxicity assay. The R code necessary to reproduce the results is given.
Subject
Medical Laboratory Technology,Toxicology,General Biochemistry, Genetics and Molecular Biology,General Medicine