Author:
Choby Garret,Trojanowski Andrzej,Johnson Samuel,Goldenberg David
Abstract
We describe a novel technique using double electrode pads for intraoperative neuromonitoring of the recurrent laryngeal nerve. This is a novel setup of intraoperative neuromonitoring that helps protect against faulty readings due to tube migration during surgery and allows for the recording of “side-specific” recurrent laryngeal nerve information. This method has not been previously described in the literature. This original approach to neuromonitoring helps guard against tube migration, simplifies the intubation process, allows for the acquisition of side-specific information, and is less expensive than use of commercially available integrated endotracheal tubes.
Subject
General Medicine,Otorhinolaryngology
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. Nerve Monitoring and Medical Malpractice;Intraoperative Cranial Nerve Monitoring in Otolaryngology-Head and Neck Surgery;2022