Contour Map Point Distribution and Surgeon Experience Level Affect Accuracy of Surgical Navigation in a Pilot Study

Author:

Talmadge Jason1ORCID,Jiang Zi Yang2ORCID,Zebda Denna A.2,Yao William C.2ORCID,Luong Amber U.23,Citardi Martin J.2

Affiliation:

1. Department of Otolaryngology & Communication Sciences, The Medical College of Wisconsin, Milwaukee, WI, USA

2. Department of Otorhinolaryngology—Head and Neck Surgery, McGovern Medical School, The University of Texas Health Science Center at Houston, Houston, TX, USA

3. Center for Immunology and Autoimmune Diseases, Institute of Molecular Medicine, McGovern Medical School, The University of Texas Health Science Center at Houston, Houston, TX, USA

Abstract

Background: Reliable use of surgical navigation depends upon the registration process. The gold standard is paired-point registration with bone-anchored fiducials, but contour-map registration is more practical. Surgeons may employ variable contour maps and less experienced team members often perform this critical step. The impact of these practices on target registration error (TRE) is not well-studied. Methods: A dry lab set-up consisting of a navigation system (Fusion ENT, Medtronic, Jacksonville, FL) and a sinus phantom with 2 mm radiopaque spheres in the sphenoid and ethmoid regions was developed. A CT (0.625 mm slice thickness) was obtained. Registration was performed with a contour-based protocol. Accuracy was determined using the software’s distance measurement tool. Registration was performed with narrow-field (NF; forehead points medial to the mid-pupillary line) and wide field (WF; entire forehead) contour maps. An experienced rhinologist and a resident surgeon performed each registration in triplicate and TRE at the sphenoid and ethmoid markers was measured in triplicate. Results: WF mapping had a lower TRE than NF (1.09 mm [95% Confidence Interval (CI) 0.96-1.22] vs 1.68 mm [95% CI 1.50-1.86]). The experienced surgeon had a lower TRE compared to the resident (1.21 mm [95% CI 1.08-1.34] vs 1.54 mm [95% CI 1.35-1.74]). Conclusions: In this navigation model, wide field mapping offers better accuracy than narrow-field mapping, and an experienced surgeon seemed to achieve better accuracy than a resident surgeon. These observations have potential implications for the use of this technology in the operating room.

Publisher

SAGE Publications

Subject

General Medicine,Otorhinolaryngology

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. The Effect of Fiducial Marker Number and Configuration on Registration Error in Dynamic Implant Surgery;The International Journal of Oral & Maxillofacial Implants;2023-07

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3