Abstract
In a patient with a multichannel cochlear implant, it was possible to demonstrate stapedius reflex contraction to intracochlear electrical stimulation. Using a standard immittance measurement technique, characteristics of the electrically evoked reflex were compared to analogous characteristics of the acoustically evoked reflex. Latency-intensity functions were similar for the two modes of excitation, but reflex waveform morphology and amplitude growth functions were different.
Subject
General Medicine,Otorhinolaryngology
Cited by
45 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献