Author:
Clifford A. R.,Gibson W. P. R.
Abstract
Twenty-three human temporal bones were dissected to study the anatomy of the round window with respect to cochlear implantation. Information was also gathered concerning the passage of a 22-channel electrode array along the cochlear duct by using cochlear surface dissection preparations. The insertion length and any insertion trauma were noted both by observation at the time of insertion and by careful analysis of video recordings. The crista semilunaris did not interfere with smooth electrode insertion. The recommendation is made that enlarging the round window anteroinferiorly is unnecessary and potentially hazardous in most cases.
Subject
General Medicine,Otorhinolaryngology
Cited by
14 articles.
订阅此论文施引文献