Abstract
Since 1980 we have used middle ear implants of aluminum oxide ceramic in some reconstructions of the ossicular chain. We had opportunities to examine 108 ears (91 patients); the longest period of postoperative observation was 4 years. Follow-up examinations showed that the rates of extrusion and liability to extrusion can be reduced by interposing a small disk of autologous cartilage and perichondrium. Moreover, interposing cartilage tends to improve audiometric results. On the basis of our study, we strongly recommend the interposition of a disk of autologous cartilage and perichondrium between the middle ear implant and eardrum when an Al2O3 implant is used.