Field strength difference in extent of artifacts induced by CERTAS Plus valves in patients with idiopathic normal pressure hydrocephalus

Author:

Camerucci Emanuele1ORCID,Elder Benjamin D23,Shu Yunhong1,Messina Steven A1,Gunter Jeffrey L1,Graff-Radford Jonathan4,Jones David T4,Botha Hugo4,Cutsforth-Gregory Jeremy K4ORCID,Jack Clifford R1,Huston John1,Cogswell Petrice M1

Affiliation:

1. Department of Radiology, Mayo Clinic, Rochester, MN, USA

2. Department of Neurologic Surgery, Mayo Clinic, Rochester, MN, USA

3. Department of Physiology and Biomedical Engineering, Mayo Clinic, Rochester, MN, USA

4. Department of Neurology, Mayo Clinic, Rochester, MN, USA

Abstract

Background and purpose : Post-shunt MRI is usually performed at 1.5T under the general assumption that shunt-related susceptibility artifacts would be greater at higher field strengths. Purpose The purpose is to show that imaging post-shunt idiopathic normal pressure hydrocephalus (iNPH) patients at 3T is feasible and with reduced artifacts as compared to 1.5T. Methods We manually measured transverse dimensions of artifact at the levels of lateral ventricles, cerebral aqueduct, and cerebellar hemisphere. Areas/volumes of artifacts were calculated assuming an elliptic/ellipsoid shape. Relative extent of shunt-related artifact between field strengths was rated by 3 readers on a 5-point Likert scale. A Wilcoxon Signed Rank Test was used to compare artifact at 1.5T vs 3T for each sequence, with a significance level set at p < 0.05. Results Artifact areas were calculated in 22 iNPH patients; artifacts were on average smaller at 3T vs 1.5T on MPRAGE, DWI, and GRE sequences. On T2 FLAIR and T2 FSE, artifacts at 3T were larger than 1.5T. On the qualitative analysis, artifact effects were less at 3T vs 1.5T on DWI, greater at 3T on T2 FSE, and had mixed results on GRE. Conclusion Our results indicate feasibility of post-shunt imaging with the CERTAS Plus valve at 3T based on shunt-related artifact that is less than or equal in extent to that on 1.5T on most standard clinical imaging sequences. Our findings, corroborated by the qualitative image review, suggest that dedicated clinical imaging sequences for devices may allow for reduction in artifact extent at both 1.5T and 3T.

Publisher

SAGE Publications

Subject

Neurology (clinical),Radiology, Nuclear Medicine and imaging,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3