Evaluating the Potential of a New Low-Profile Urinary Catheter in Preventing Catheter-Associated Urinary Tract Infections: A Prospective Randomized Blinded Clinical Trial

Author:

Ramezani Farahnaz1,Khatiban Mahnaz2ORCID,Rahimbashar Farshid13,Soltanian Ali Reza4,Mousavi-bahar Seyed Habibollah15,Elyasi Ensieh1

Affiliation:

1. Besat Specialized and Sub-specialized Hospital, Hamadan University of Medical Sciences, Hamadan, Iran

2. Mother and Child Care Research Center, Department of Medical-Surgical Nursing, Department of Ethics Education in Medical Sciences, School of Nursing and Midwifery, Hamadan University of Medical Sciences, Hamadan, Iran

3. Department of Anesthesiology, School of Medicine, Hamadan University of Medical Sciences, Hamadan, Iran

4. Modeling of Noncommunicable Disease Research Center, Department of Biostatistics and Epidemiology, School of Public Health, Hamadan University of Medical Sciences, Hamadan, Iran

5. Urology & Nephrology Research Center, Hamadan University of Medical Sciences, Hamadan, Iran

Abstract

Aim To investigate the efficacy of a new low-profile catheter on incidence of the catheter-associated urinary tract infections (CAUTI) in comatose patients admitted to the intensive care unit. Background Catheter-induced urothelial injury is a key component in the development of urinary tract infections in catheterized patients. Methods In this prospective randomized blinded clinical trial, 80 patients requiring indwelling urinary catheterization were equally randomized to either the standard Foley catheter (control) or the low-profile catheter (experimental) group. The signs of urinary tract infection for comatose patients were considered (ie, ≥105 of colony-forming unit/milliliter of urine, hematuria, serum leukocytes, and body temperature) and recorded at baseline and on days 3 and 5 after catheterization. The analysis of covariance was applied by the SPSS-20 software at a 95% confidence level. Results An increasing proportion of patients with elevated urinary colony counts were seen in the Foley catheter group compared with the low-profile catheter group (12.5% vs 5%). However, there were no between-group differences in the urinary colony counts and body temperature after controlling for antibiotic doses and fluid intake. Patients in the low-profile catheter group had significantly lower rates of hematuria and serum leukocytes than those in the Foley catheter group. Conclusion A newly designed low-profile urinary catheter has demonstrated a trend toward reducing the incidence of CAUTI in patients with indwelling urinary catheters. Further studies with larger sample sizes and follow-up are needed to confirm the benefits.

Funder

Vice Chancellor for Research and Technology, Hamadan University of Medical Sciences