Radiologic Evaluation of the Cement Mantle in the Femoral Shaft. A Comparative Study Using a Fast-Set or a Standard-Set Cement

Author:

Gallart X.1,Martínez F.1,García S.1,Riba J.1,Combalía A.1,Ballesteros J.1,Elefante F.2,Calonego G.2

Affiliation:

1. Department of Orthopedics, Hospital Clinic, University of Barcelona, Barcelona - Spain

2. Research and Development Department, Tecres Spa, Verona - Italy

Abstract

Introduction At present a wide spectrum of cement types are available for cemented total hip arthroplasty (CTHA). All types have the same objective, i.e. to prolong the life of the implant for as long as possible. The purpose of this study was to compare the radiologic cement mantle of CTHA using cement with fast or standard setting characteristics. Method A prospective comparative study of patients who underwent CTHA was performed: the first group received fast setting cement (Cemex System Fast), the second group received standard setting cement (Cemex System). A radiologic evaluation of the cement mantle was done using Barrack's classification. In order to give the study clinical relevance we assembled Barrack's classification in two groups: 1) Low risk group (Types A and B); 2) High risk group (Types C and D). According to ISO 5833:2002, the international standard was used to perform physico-chemical (polymerization temperature, setting time) and mechanical testing (compression strength, bending strength and bending modulus). Results The fast setting cement showed a shorter setting time and a higher polymerization temperature. The mechanical performances of both cements met the ISO limits. At a mean follow-up of 23 months (11–37 months) no statistically significant radiological difference was found in the radiologic cement mantle between the two groups either in the post-operative X-ray evaluation (p=0.32) or in the last follow-up (p=0.72). Statistical evaluation was done comparing the two subgroups, i.e. low risk and high risk. Conclusions The use of fast or standard setting cement does not produce any difference in the quality of the radiologic cement mantle in the short-term. Both cements comply with ISO 5833 requirements.

Publisher

SAGE Publications

Subject

Orthopedics and Sports Medicine,Surgery

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3