Abstract
The investigation of the flow field downstream of a cardiac valve prosthesis is a well established task. In particular turbulence generation is of interest if damage to blood costituents is to be assessed. Several prosthetic valve flow studies are available in literature but they generally concern large-sized prostheses. The FDA draft guidance requires the study of the maximum Reynolds number conditions for a cardiac valve model to assess the worst case in turbulence by chosing both the minimum valve diameter and a high cardiac output value as protocol set up. Within the framework of a national research project regarding the characterization of cardiovascular endoprostheses, the Laboratory of Biomedical Engineering is currently conducting an in-depth study of turbulence generated downstream of bileaflet cardiac valves. Four models of 19 mm sized bileaflet valve prostheses, namely St Jude Medical HP, Edwards Tekna, Sorin Bicarbon, and CarboMedics, were studied in aortic position. The prostheses were selected for the nominal annulus diameter reported by the manufacturers without any assessment of the valve sizing method. The hemodynamic function was investigated using a bidimensional LDA system. Results concern velocity profiles during the peak flow systolic phase, at high cardiac output regime, highlighting the different flow field features downstream of the four small-sized cardiac valves.