Affiliation:
1. Renal Unit, General Hospital, Piacenza - Italy
Abstract
Increasing the survival of patients on CAPD is related to the long-term reliability of the peritoneal access. Six silicone Tenckhoff catheters (with strip or diffuse barium sulphate inclusion) removed after 39-69 months because of the appearance of external segment fissures, were analysed by scanning electron microscopy (SEM) and infrared spectroscopy with attenuated total refractance (ATR). The extracorporeal portion of the catheters showed (by ATR) a more prominent oxidation peak on the external than the internal surface; SEM showed marks and cracks on the external surface and exfoliation and flattening of the silastic reticle on the intraluminal surface. No evidence of oxidation was found in the intra-abdominal portion of the catheters but biofilm was found. We suggest that barium sulphate may render the silastic brittle and physiological and environmental long-term factors (such as uv-rays, temperature, sweat and disinfectants) could cause oxidation and loss of physico-chemical properties, with critical aging of the silastic and loss of catheter resistance to mechanical injury.
Subject
Biomedical Engineering,Biomaterials,General Medicine,Medicine (miscellaneous),Bioengineering