Aachen-Keratoprosthesis as Temporary Implant. Case Report on First Clinical Application

Author:

Kompa S.1,Langefeld S.2,Kirchhof B.2,Brenman K.3,Schrage N.2

Affiliation:

1. Interdisciplinary Center of Clinical Research “BIOMAT”, University of Aachen

2. Department of Ophthalmology, University of Aachen - Germany

3. Department of Biomedical Engineering, University of Miami, Miami, Florida - USA

Abstract

Background The Aachen-Keratoprosthesis was designed to serve as a permanent keratoprosthesis, modeling natural corneal physical properties as closely as possible. Prior to permanent application in patients, keratoprostheses are commonly tested in animal models to assess biochemical and biomechanical compatibility. However, immune and inflammatory responses acquired through animal experimentation are difficult to extrapolate in order to develop a predictable and generalized outcome in humans. Therefore, this preliminary report includes results following a temporary implantation of the Aachen-Keratoprosthesis in a patient during vitreoretinal surgery to assess the long-term prospect of application as a permanent artificial cornea. Patient and Methods A 43 year old man presented with a ruptured right eye resulting in an opaque cornea and retinal detachment. A soft silicone rubber keratoprosthesis, the Aachen-Keratoprosthesis, was temporarily implanted. Subretinal membranes were removed and the total retinal detachment was reattached. Liquid perfluorocarbon and silicone oil were used. The temporary keratoprosthesis was replaced by a 7 mm corneal graft after completion of surgery. Results The Aachen-Keratoprosthesis was securely positioned into the trephined hole. It allowed complete visualisation of vitreous, retina and thus controlled manipulations in the vitreous cavity up to the extreme periphery. Leakage across the trephination-prosthesis interface was minimal. Conclusion: We report the first temporary application of an innovative keratoprosthesis. Its flexibility and good optical qualities allow visualisation and control of intraoperative procedures. This temporary pilot study of the Aachen KPro encourages further investigation of the Aachen KPro as a permanent replacement for a diseased cornea.

Publisher

SAGE Publications

Subject

Biomedical Engineering,Biomaterials,General Medicine,Medicine (miscellaneous),Bioengineering

Cited by 10 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3