Solutes removal characteristics at various effluent rates during different continuous renal replacement therapy modalities

Author:

Yu Wenyan1,Zhuang Feng1,Ma Shuai1,Zhu Mingli2,Ding Feng1

Affiliation:

1. Division of Nephrology, Shanghai Ninth People’s Hospital, Shanghai JiaoTong University, School of Medicine, Shanghai, China

2. Division of Nephrology, Renji Hospital, Shanghai JiaoTong University, School of Medicine, Shanghai, China

Abstract

Background:Some studies suggest the effluent as a surrogate solute removal indicator in continuous hemodialysis or hemofiltration, but the delivered clearance is frequently smaller than prescribed. This study aims at testing whether the effluent, represented by mL/kg/h, could measure solute clearance and whether increasing effluent increases clearance proportionately in continuous hemodialysis or hemofiltration.Methods:Patients treated with continuous renal replacement therapy for various diagnoses were included. The range of dialysate flow rate or substitution fluid flow rate was 1–5 L/h; solutes in the effluent and in the plasma entering the filter were measured, and the ratio of solutes in the effluent and in the plasma entering the filter and the clearance of blood urea nitrogen, creatinine, phosphate, and β2-microglobulin were calculated.Results:The ratio of solutes in the effluent and in the plasma entering the filter showed a decreasing trend with increased dialysate flow rate or substitution fluid flow rate ( p  < 0.05), but solute clearance showed an increasing trend. The increase in solute clearance was less than expected from the increased effluent ( p < 0.01), and actual delivered clearance was always below the corresponding prescribed clearance ( p < 0.001).Conclusion:With increasing prescribed clearance of continuous renal replacement therapy, effluent rate overestimated the delivered clearance.

Funder

national natural science foundation of china

science and technology commission of shanghai municipality

Shanghai Pudong New Area Program

Shanghai Jiaotong University Interdisciplinary Key Program

Shanghai Ninth People’s Hospital Clinical Research Program

Publisher

SAGE Publications

Subject

Biomedical Engineering,Biomaterials,General Medicine,Medicine (miscellaneous),Bioengineering

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Precision renal replacement therapy;Current Opinion in Critical Care;2020-09-29

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3