Affiliation:
1. Department of Biomedical Engineering, Seoul National University, Seoul - Korea
Abstract
A new automatic cardiac output control algorithm for an implantable electromechanical total artificial heart (TAH) was developed based on the analysis of motor current waveform without using any transducer. The basic control requirements of an artificial heart can be described in terms of three features: preload sensitivity, afterload insensivity, and balanced ventricular output. In previous studies, transducers were used to acquire information on the hemodynamic states for automatic cardiac output control. However, such a control system has reliability problems with the sensors. We proposed a novel sensorless automatic cardiac output control algorithm (ACOCA) providing adequate cardiac output to the time-varying physiological demand without causing right atrial collapse, which is one of the critical problems in an active filling device. In vitro tests were performed on a mock circulatory system to assess the performance of the developed algorithm and the results show that the new algorithm satisfied the basic control requirements of the cardiac output response.
Subject
Biomedical Engineering,Biomaterials,General Medicine,Medicine (miscellaneous),Bioengineering