Affiliation:
1. Helmholtz-Institute for Biomedical Engineering at the RWTH Aachen, Aachen - Germany
Abstract
Mock loops are an important tool for in vitro investigations of artificial blood pumps. The simple windkessel, throttle, and atrium principle was used for the mock loop design presented. The components of the systemic and the pulmonary mock loop were designed according to calculated numerical simulation parameters. The loops offer a compact design and simple handling. For simulating biventricular assist or total artificial heart (TAH), both loops can be coupled correspondingly. The numerical simulation and the first results with the loops show very good similarity to physiological data of systemic and pulmonary circulation. The measurements of pump characteristics are significant for quantitative comparison of different pump sizes and types, or driving systems.
Subject
Biomedical Engineering,Biomaterials,General Medicine,Medicine (miscellaneous),Bioengineering
Cited by
44 articles.
订阅此论文施引文献