Residual Blood Loss in Single Use Dialyzers: Effect of Different Membranes and Flux

Author:

Kalocheretis P.1,Vlamis I.1,Belesi C.2,Makriniotou I.1,Zerbala S.1,Savidou E.2,Zorbas S.1,Arvanitis N.1,Iatrou C.1

Affiliation:

1. Center for Nephrology “G. Papadakis”, General Hospital of Nikea, Pireaus - Greece

2. Division of Hematology, General Hospital of Nikea, Pireaus - Greece

Abstract

Aim/Background The aim of this study was to evaluate the residual blood loss in new type single use dialyzers under the usually prevailing conditions during hemodialysis and to investigate whether or not this loss is dependent on dialyzer membrane composition or flux characteristics. Materials and Methods In 158 hemodialysis (HD) patients, 158 single used dialyzers were studied in corresponding HD sessions. 52/158 dialyzers were made from modified cellulose (acetate, CA or triacetate, CTA) membrane and 106 from synthetic ones (58 with ethyl-vinyl-alcohol (EVAL), 48 with polyacrilonitrile (AN69)). Of those dialyzers 85/158 (58 EVAL+27CA) were low flux (LF) while the other 73 were high flux (HF). Patients underwent 4 hour HD sessions and at the end of the session blood was drawn for the measurement of hematocrit (Ht) and hemoglobin (Hb). Additionally, after the end of dialysis the used dialyzers were rinsed with 1000mL of 0.05% NH3 solution in distilled water. The wash was collected and subsequently Hb was measured using the benzidine method. From the volume of the solution and its concentration of Hb, total Hb of the solution was measured and blood loss in terms of red blood cell (RBC) volume was estimated by the use of the formula: RBC (mL) = Total Hb (g) in the solution x patient's Ht (ml/dL) / Patient's Hb (g/dL). For results to be comparable between dialyzers, RBC volume/m2 of dialyzer membrane was expressed. In 5/158 patients blood loss was also estimated in 6 consecutive HD sessions using the same type of dialyzer. Results For the sum of the dialyzers, blood loss / dialyzer in terms of RBC volume, expressed as median (range), was 0.978 mL (0.01–23.9). There was statistically significant (p<0.001 or p<0.05) higher blood loss with the use of AN69 dialyzer than with the other three. RBCHF>RBCLF (p<0.001) constrained the first group of patients to use a 6% higher dosage of ferrum and 3.5% higher dosage of erythropoietin than the other group to achieve the optimal hemoglobulin values. No difference existed regarding RBC between CA, CTA and EVAL membranes. RBC measured in the small group of 5/158 patients for 6 consecutive HD sessions with the same dialyzer showed a wide range of RBC loss indicating an effect of the human factor. Conclusions Blood loss during HD sessions due to residual blood cell volume inside dialyzers is usually slight using new type single use dialyzers but, sometimes, it can be significant and may contribute to the development or deterioration of preexisting iron deficiency anemia. The results of this study indicated that this loss can be attributed to the membrane composition of the dialyzer or to the human factor and has nothing to do with the ultrafiltration coefficient of the dialyzer.

Publisher

SAGE Publications

Subject

Biomedical Engineering,Biomaterials,General Medicine,Medicine (miscellaneous),Bioengineering

Cited by 12 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3