A total of 15 bileaflet mechanical heart valves were studied in a pulse duplicator at the Helmholtz Institute (Aachen, Germany) under conditions approximating first, a physiological pressure curve and subsequently, a sinusoidal pressure curve. In this study Edwards-Duromedics valves of the modified specification were compared with the earlier version of the Edwards-Duromedics valve as well as with St. Jude Medical valves. Each valve was tested at a series of nine (9) conditions. At each condition, without altering the valve installation or the systemic conditions, each valve was filmed by two separate video systems: the Helmholtz Institute strobe light system and a high speed video recording system. All data, as recorded by each system, was then independently analyzed by both of the two contributing groups and subsequently compared. In this manner, it was possible to objectively verify not only the consistency of the data obtained, but to also determine the relative reliability of the methods for cavitation threshold detection.