Errors in Fluid Balance with Pump Control of Continuous Hemodialysis

Author:

Roberts M.1,Winney R.J.2

Affiliation:

1. Nephrology Department, Sepulveda VA Medical Center, Sepulveda, CA - USA

2. Medical Renal Unit, Royal Infirmary, Edinburgh - UK

Abstract

The use of pumps both proximal and distal to the dialyzer during continuous hemodialysis provides control of dialysate and ultrafiltration flow rates, thereby reducing nursing time. However, we had noted unexpected severe extracellular fluid depletion suggesting that errors in pump delivery may be responsible. We measured in vitro the operation of various pumps under conditions similar to continuous hemodialysis. Fluid delivery of peristaltic and roller pumps varied with how the tubing set was inserted in the pump. Piston and peristaltic pumps with dedicated pump segments were more accurate. Pumps should be calibrated and tested under conditions simulating continuous hemodialysis prior to in vivo use.