Longer stitch interval in the Krackow stitch for tendon graft fixation leads to poorer biomechanical property

Author:

Hong Chih-Kai1,Lin Cheng-Li1,Kuan Fa-Chuan1,Wang Ping-Hui2,Yeh Ming-Long34,Su Wei-Ren145

Affiliation:

1. Department of Orthopaedic Surgery, National Cheng Kung University Hospital, College of Medicine, National Cheng Kung University, Tainan, Taiwan

2. Department of Orthopedics, Chi-Mei Medical Center, Tainan, Taiwan

3. Department of Biomedical Engineering, National Cheng Kung University, Tainan, Taiwan

4. Medical Device Innovation Center, National Cheng Kung University, Tainan, Taiwan

5. Medical Device R&D Core Laboratory, National Cheng Kung University Hospital, Tainan, Taiwan

Abstract

Purpose: The purpose of this study was to analyze the effects of different intervals between stitch throws on tendon graft fixation with the Krackow stitch. Methods: Forty-four porcine flexor profundus tendons were randomly divided into four groups of 11 specimens each. The Krackow stitch with various stitch intervals (2.5, 5.0, 7.5, and 10.0 mm) were evaluated, and named the K-2.5, K-5.0, K-7.5, and K-10.0 groups, respectively. A braided nonabsorbable suture was used to complete each suture-tendon construct. All specimens were pretensioned to 100 N for three cycles, cyclically loaded from 50 to 200 N for 200 cycles, and then finally loaded to failure. Elongation after cyclic loading, ultimate load to failure, and the mode of failure were recorded. Results: There were significant differences in elongation after cyclic loading among the K-2.5 (31% ± 5%), K-5.0 (32% ± 4%), K-7.5 (34% ± 5%), and K-10.0 (41% ± 8%) groups ( p = 0.004); the post hoc analysis showed significantly smaller values in the K-2.5 and K-5.0 groups than in the K-10.0 group ( p = 0.002 and 0.003, respectively). The stitch interval was correlated with elongation after cyclic loading ( r = 0.52, p < 0.001). Ultimate loads to failure and cross-sectional area were not significantly different across the four groups. Conclusion: The Krackow stitch with stitch intervals of 2.5 and 5.0 mm had significantly smaller elongation after cyclic loading than with an interval of 10.0 mm in this porcine biomechanical study. The stitch interval was moderately correlated with elongation after cyclic loading.

Publisher

SAGE Publications

Subject

Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3