Affiliation:
1. Department of Laboratory Medicine, Children's National Medical Center, 111 Michigan Avenue, N.W., Washington, D.C. 20010
2. Departments of Child Health and Development and Pathology, George Washington School of Medicine and Children's National Medical Center, Washington, DC
Abstract
Three desktop chemistry analyzers most frequently utilized in physicians' offices are compared for performance and precision: the DT60 (Kodak), Vision™ (Abbott), and Reflotron® (Boehringer-Mannheim). The manufacturers' instructions were followed for all the reagents, calibrators and controls provided with each analyzer used. Precision, stability, dynamic range, interferences, and comparison of analyses of 50-100 patient samples with standard laboratory instruments were evaluated. Within-run precision on all three instruments was acceptable, the coefficient of variation (% CV) ranging from 0.8-5.4% and run-to-run precision CV of 1.1-7.3%. Within-run precision for glucose, blood urea, nitrogen, cholesterol and sodium for an experienced technologist, a clerk without any laboratory experience and a physician with no recent laboratory training were compared: the total CV varied between 1.1 and 6.3%. There are differences in the way bilirubin and hemoglobin interfere with the chemistries on the analyzers. Instrument size, available analyses, reagent cost and ease of operation were also assessed. The data suggest that all three analyzers performed well and would be suitable for the pediatric office setting.
Subject
Pediatrics, Perinatology and Child Health
Cited by
10 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献