Affiliation:
1. Department of Vascular Surgery, Scientific Institute H. San Raffaele, Vita Salute San Raffaele University, Milan, Italy
Abstract
Purpose: To present a Supera stent fracture following treatment of popliteal artery stenosis. Case Report: A 60-year-old man previously treated with angioplasty/stenting of a popliteal artery lesion with a Supera stent presented at 8 months with in-stent occlusion that proved to be secondary to stent fracture. Conclusion: Supera interwoven nitinol stents, despite having higher radial force and conformability than classic tube stents, are not free from fracture. Moreover, due to their peculiar structure, a break in these stents results in complete loss of integrity and consequent device collapse, thus jeopardizing endovascular recanalization and relining.
Subject
Cardiology and Cardiovascular Medicine,Radiology Nuclear Medicine and imaging,Surgery
Cited by
21 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献