Preliminary Clinical Results and Mechanical Behavior of a New Double-Layer Carotid Stent

Author:

Wissgott Christian1,Schmidt Wolfram2,Brandt Christoph2,Behrens Peter2,Andresen Reimer1

Affiliation:

1. Institute of Diagnostic and Interventional Radiology/Neuroradiology, Westkuestenklinikum Heide–Academic Teaching Hospital of the Universities of Kiel, Luebeck and Hamburg, Heide, Germany

2. Institute for Biomedical Engineering, University Medicine, University of Rostock, Germany

Abstract

Purpose: To evaluate the in vitro mechanical and clinical implant behavior of a next-generation double-layer stent designed for the carotid artery. Methods: The new double-layer CASPER-RX stent was implanted in 12 patients (median age 69 years; 8 men) with high-grade symptomatic internal carotid artery stenoses (mean 82%). In the in vitro experiments, the CASPER-RX stent (8-×40-mm model) was investigated with respect to its radial force on expansion and the bending stiffness of the stent system and of the stent in its expanded state, as well as the collapse pressure in a thin, flexible sleeve. The wall adaptation of the expanded stents was assessed by fluoroscopy after release in a step and curve model. Results: Technical success was achieved in all patients without complications; there was no peri- or postinterventional stroke and no stroke or restenosis after 6 months. In the experimental studies, the bending stiffness of the stent on the delivery system (154.9 N mm2) was significantly lower than when expanded in a 7-mm flexible tube (467.4 N mm2). The radial force on expansion of the stent to 7 mm was low (0.011 N/mm). The collapse pressure was relatively high (0.56 bar) as a result of the stent’s particular stent structure. The stent exhibited significant foreshortening of 27.6%. The conformability to the wall in the step model was relatively smooth; in the curve model, straightening occurred with consecutive slight stenosis. Conclusion: The first clinical results showed a safe implantation behavior without the occurrence of any ischemia. The structure of the new CASPER-RX stent creates an acceptable flexibility, low radial force, and high collapse pressure. The large foreshortening during implantation should be considered as well as the higher bending stiffness, especially when used in elongated carotid arteries.

Publisher

SAGE Publications

Subject

Cardiology and Cardiovascular Medicine,Radiology Nuclear Medicine and imaging,Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3