Comparison of Angiographic Dissection Classification Systems in the Femoropopliteal Arteries Using IVUS Validation and Reliability Testing

Author:

Allan Richard Barry12ORCID,Wise Nadia Clare12,Wong Yew Toh12,Delaney Christopher Luke12

Affiliation:

1. Department of Vascular and Endovascular Surgery, Flinders Medical Centre, Bedford Park, SA, Australia

2. College of Medicine and Public Health, Flinders University, Adelaide, SA, Australia

Abstract

Purpose: Objective assessment of dissection severity is difficult. Recognition of this has led to the creation of classification systems. This study investigated the performance of the National Heart Lung and Blood Institute (NHLBI) and Kobayashi systems at differentiating severity of femoropopliteal dissection using intravascular ultrasound (IVUS) as the reference standard. Comparison between the 2 systems and the inter- and intra-observer reliability were also investigated. Materials and Methods: Angiographic and IVUS imaging was assessed in 51 cases sourced from a RCT investigating the use of IVUS in femoropopliteal disease. A total of 2 readers independently scored the angiography images according to NHLBI and Kobayashi dissection classification systems and a consensus score was obtained for each system in each case. The NHLBI classification was condensed into 3 grades of dissection to allow comparison between systems. Dissection length, dissection arc, minimum lumen area, and lumen area stenosis were obtained from the IVUS imaging. IVUS parameters were compared between grades of severity for both systems. Agreement in grading between the systems was assessed and IVUS parameters for each level of dissection severity were compared between systems. Inter and intra-observer agreement tested for each system. Results: Dissection was present on IVUS in 92.2% (47/51) of cases and angiography identified 78.7% (37/47) of these. No difference was present in any IVUS parameters between mild and severe dissections with either classification system. No difference in IVUS findings was present for the same grades of dissection between systems. The 2 systems agreed on severity grade in 47 of 51 cases. The inter-observer agreement was for NHLBI was k=0.549 and k=0.627 for Kobayashi. Intra-observer agreement for NHLBI was k=0.633 and k=0.633 and for Kobayashi was k=0.657 and k=0.297. Conclusion: The lack of difference in IVUS parameters between mild and severe dissection for the NHLBI and Kobayashi systems raises doubts about their ability to effectively differentiate dissection severity. Weak to moderate reliability suggests that variability in interpreting dissection may be higher than acceptable. IVUS imaging is more sensitive for detecting dissection than angiography and research is required to establish the value of adding IVUS to dissection classification systems.

Funder

National Heart Foundation of Australia

Publisher

SAGE Publications

Subject

Cardiology and Cardiovascular Medicine,Radiology Nuclear Medicine and imaging,Surgery

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3