Air Embolism During TEVAR: An Additional Flush Port on the Delivery System Pusher Significantly Reduces the Amount of Air Released During Deployment of a Thoracic Stent-Graft in an Experimental Setting

Author:

Rohlffs Fiona1ORCID,Tsilimparis Nikolaos1,Trepte Constantin2,Kratzberg Jarin3,Mogensen John3,Debus E. Sebastian1,Kölbel Tilo1

Affiliation:

1. German Aortic Center, Department of Vascular Medicine, University Heart Center, University Hospital Hamburg-Eppendorf, Hamburg, Germany

2. Department of Anesthesiology, University Hospital Hamburg-Eppendorf, Hamburg, Germany

3. Cook Medical, Bjaeverskov, Denmark

Abstract

Purpose: To investigate the influence of (1) an additional side port for flushing the hollow pusher in Zenith thoracic stent-graft delivery systems and (2) additional carbon dioxide flushing on the amount of air released during stent-graft deployment. Methods: Twenty thoracic stent-grafts with an additional flush port to fill the hollow pusher were separated into 2 equal groups (C and D). Both groups were flushed with 20 mL of normal saline through the extra side port connected to the pusher and with 60 mL of saline through the regular flushing port. One group of grafts (group D) was additionally flushed with carbon dioxide through the regular flushing port prior to saline. All grafts were deployed into a curved plastic pipe attached to the bottom of a water-filled container. The released gas was recorded and measured using a calibrated setup. To evaluate the influence of the extra side port irrespective of the carbon dioxide flushing technique, group C was compared with a previously published reference group A without an extra side port that was flushed with the standard 60 mL of saline. Results: Volumes of gas were released in various amounts from the stent-grafts during deployment. The average amount of released gas was 0.51 mL in group C and 0.07 mL in group D (p<0.001). The mean amount of gas from group C samples (0.51 mL) was also significantly lower (p=0.002) compared with the reference group (0.79 mL). Conclusion: Thoracic endografts release air during deployment. Reducing the air-filled space inside the pusher of the catheter assembly using an additional side port can significantly reduce the amount of released air. Using the extra side port in combination with the carbon dioxide flushing technique reduces gas release further to small volumes. In a clinical setting this could be a promising approach to lower the risk of air embolism and stroke during thoracic endovascular aortic repair.

Publisher

SAGE Publications

Subject

Cardiology and Cardiovascular Medicine,Radiology, Nuclear Medicine and imaging,Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3