Pulsatile Changes of the Aortic Diameter May Be Irrelevant Regarding Endograft Sizing in Patients With Aortic Disease

Author:

Fontanini Daniele Mariastefano12ORCID,Huber Máté1,Vecsey-Nagy Milán2,Borzsák Sarolta12,Csőre Judit1,Sótonyi Péter23,Csobay-Novák Csaba12ORCID

Affiliation:

1. Department of Interventional Radiology, Semmelweis University, Budapest, Hungary

2. Semmelweis Aortic Center, Heart and Vascular Center, Semmelweis University, Budapest, Hungary

3. Department of Vascular and Endovascular Surgery, Semmelweis University, Budapest, Hungary

Abstract

Purpose: Endovascular aortic repair has become the preferred elective treatment of infrarenal aortic aneurysms. Aortic pulsatility may pose problems regarding endograft sizing. The aims of this study are to determine the aortic pulsatility in patients with aortic disease and to evaluate the effect of pulsatility on the growth of aneurysms. Materials and Methods: In this retrospective study, analyses of computed tomography angiography (CTA) images of 31 patients under conservative treatment for small abdominal aortic aneurysms were performed. Reconstructions of the raw electrocardiography (ECG) gated dataset at 30% and 90% of the R-R cycle were used. After lumen segmentation, total aortic cross-sectional area was measured in diastole and systole in the following zones: Z0, Z3, Z5, Z6, Z8, and Z9. Effective diameters (EDs) were calculated from the systolic (EDsys) and diastolic (EDdia) cross-sectional areas to determine absolute (EDsys – EDdia, mm) and relative pulsatility [(EDsys – EDdia) / EDdia, %]. Diameter of the aneurysms was measured on baseline images and the last preoperative follow-up study of each patient. Results: A total of 806 measurements were completed, 24 pulsatility and 2 growth measurements per patient. The mean pulsatility values at each point were as follows: Z0: 0.7±0.8 mm, Z3: 1.0±0.6 mm, Z5: 1.0±0.6 mm, Z6: 0.8±0.7 mm, Z8: 0.7±1.0 mm, Z9: 0.9±0.9 mm. Follow-up time was 5.5±2.2 years during which a growth of 13.42±9.09 mm (2.54±1.55 mm yearly) was observed. No correlation was found between pulsatility values and growth rate of the aneurysms. Conclusion: The pulsatility of the aorta is in a submillimetric range for the vast majority of patients with aortic disease, thus probably not relevant regarding endograft sizing. Pulsatility of the ascending aorta is smaller than that of the descending segment, making an additional oversize of a Z0 implantation questionable. Clinical Impact Endovascular aortic repair reqiures precise preoperative planning. Pulsatile changes of the aortic diameter may pose issues regarding endograft sizing. In our retrospective single-centre study, aortic pulsatility of patients with AAA was measured on ECG gated CTA images. Pulsatility values reached a maximum at the descending aorta, however absolute pulsatility values did not exceed 1 mm at any point along the aorta. Therefore, significance of aortic pulsatility regarding the sizing of EVAR prostheses is questionable. Correlation between pulsatility and AAA growth was not found.

Funder

Ministry for Innovation and Technology in Hungary

National Research, Development and Innovation Fund of Hungary

Publisher

SAGE Publications

Subject

Cardiology and Cardiovascular Medicine,Radiology, Nuclear Medicine and imaging,Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3