In Situ Laser Fenestration Technique: Bench-Testing of Aortic Endograft to Guide Clinical Practice

Author:

Grima Matthew Joe123ORCID,Wanhainen Anders14ORCID,Lindström David1ORCID

Affiliation:

1. Department of Surgical Sciences, Section of Vascular Surgery, Uppsala University, Uppsala, Sweden

2. Faculty of Medicine and Surgery, University of Malta, L-iMsida, Malta

3. Department of General Surgery, Vascular Unit, Mater Dei Hospital, L-iMsida, Malta

4. Department of Surgical and Perioperative Sciences, Surgery, Umeå University, Umeå, Sweden

Abstract

Purpose: In situ laser fenestration (ISLF) is a recently introduced technology that offers the potential to perform total endovascular treatment of aortic arch and thoracoabdominal aortic pathologies in the acute setting. This experiment’s aim was to assess ISLF in some currently common aortic endografts and bridging stent-grafts. Materials and Methods: Three different aortic endografts were evaluated: (1) Zenith Alpha, (2) Zenith TX2, and (3) Conformable GORE TAG. Each endograft was submerged in 37°C saline to create fenestrations using the 308 nm CVX-300 Excimer Laser System fitted with a 2.3 mm diameter Turbo-Elite laser atherectomy catheter compatible with a 0.018″ guidewire. Three different 8 mm bridging stent-grafts were evaluated: (1) BeGraft peripheral, (2) BeGraft peripheral plus, and (3) GORE VIABAHN VBX Balloon Expandable. All bridging stent-grafts were deployed and exposed to different balloon sizes and pressures. The ISLFs and bridging stent-grafts were then evaluated for any tears, stenoses, and seal. Results: A laser fenestration was consistently rapidly obtained in the Zenith Alpha and the Zenith TX2 endografts while it proved difficult to achieve a timely fenestration in the C-TAG. No fabric tears were noted in the Zenith Alpha and Zenith TX2 when inflating Armada (Abbott) 8 mm balloon in the fenestrations with pressures up to 15 atmospheres (rated burst pressure) nor when flaring bridging stent-grafts with balloons up to 12 mm in diameter at 10 atmospheres, while major tears were frequently noted in the C-TAG when the Armada 8 mm balloons were inflated. BeGraft Peripheral and BeGraft Peripheral Plus were all firmly attached to the fenestrations showing good seal on manual testing, while every sixth VBX bridging stent-graft displayed poorer attachment to the fenestration before dilatation at high pressure. Commonly, significant stenoses remained in the bridging stent-grafts after dilatation at nominal pressure, which could only be eradicated with high-pressure balloons. Conclusion: In this limited bench-test, Dacron endografts responded well to the ISLF technology. Satisfactory deployment of the bridging stent was noted only after inflation and/or flaring with high-pressure balloons. Further work with different types of commercially-available bridging stent-grafts and endografts to assess the durability of in situ fenestration (ISF) and bridging stents in ISF is recommended. Clinical Impact This report on experimental in situ laser fenestration provide important insights for clinicians considering using in situ laser fenestration of aortic stentgrafts in vivo. In particular, different laser settings were tested together with a selection of aortic stentgrafts. Also, the target pressure needed in PTA balloons to dilate the fenestrations and any subsequent tears in the fabric were noted. This was followed by deployment of assorted balloon-expandable stentgrafts with estimation of residual stenosis and seal.

Publisher

SAGE Publications

Subject

Cardiology and Cardiovascular Medicine,Radiology, Nuclear Medicine and imaging,Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3