In Vitro Quantification of Gutter Formation and Chimney Graft Compression in Chimney EVAR Stent-Graft Configurations Using Electrocardiography-Gated Computed Tomography

Author:

Overeem Simon P.12ORCID,Donselaar Esmé J.3,Boersen Jorrit T.123,Groot Jebbink Erik13,Slump Cornelis H.1,de Vries Jean-Paul P. M.2,Reijnen Michel M. P. J.3ORCID

Affiliation:

1. MIRA Institute for Biomedical Technology and Technical Medicine, University of Twente, Enschede, the Netherlands

2. Department of Vascular Surgery, St Antonius Hospital, Nieuwegein, Utrecht, the Netherlands

3. Department of Vascular Surgery, Rijnstate Hospital, Arnhem, the Netherlands

Abstract

Purpose: To assess the dynamic behavior of chimney grafts during the cardiac cycle. Methods: Three chimney endovascular aneurysm repair (EVAR) stent-graft configurations (Endurant and Advanta V12, Endurant and Viabahn, and Endurant and BeGraft) were placed in silicone aneurysm models and subjected to physiologic flow. Electrocardiography (ECG)-gated contrast-enhanced computed tomography was used to visualize geometric changes during the cardiac cycle. Endograft and chimney graft surface, gutter volume, chimney graft angulation over the center lumen line, and the D-ratio (the ratio between the lengths of the major and minor axes) were independently assessed by 2 observers at 10 time points in the cardiac cycle. Results: Both gutter volumes and chimney graft geometry changed significantly during the cardiac cycle in all 3 configurations (p<0.001). Gutters and endoleaks were observed in all configurations. The largest gutter volume (232.8 mm3) and change in volume (20.7 mm3) between systole and diastole were observed in the Endurant-Advanta configuration. These values were 2.7- and 3.0-fold higher, respectively, compared to the Endurant-Viabahn configuration and 1.7- and 1.6-fold higher as observed in the Endurant-BeGraft configuration. The Endurant-Viabahn configuration had the highest D-ratio (right, 1.26–1.35; left, 1.33–1.48), while the Endurant-BeGraft configuration had the lowest (right, 1.11–1.17; left, 1.08–1.15). Assessment of the interobserver variability showed a high correlation (intraclass correlation >0.935) between measurements. Conclusion: Gutter volumes and stent compression are dynamic phenomena that reshape during the cardiac cycle. Compelling differences were observed during the cardiac cycle in all configurations, with the self-expanding (Endurant–Viabahn) chimney EVAR configurations having smaller gutters and less variation in gutter volume during the cardiac cycle yet more stent compression without affecting the chimney graft surface.

Publisher

SAGE Publications

Subject

Cardiology and Cardiovascular Medicine,Radiology Nuclear Medicine and imaging,Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3