Affiliation:
1. Department of Radiology, Christian-Albrechts-University, Kiel, Germany
Abstract
Purpose: To describe a method for dealing with balloon rupture during stent deployment. Technique: A 10-mL Luer-Lock syringe containing contrast material and heparinized saline is used to re-expand a balloon ruptured during stent deployment, permitting maximum balloon expansion and successful initial stent-wall apposition. No adjunctive use of probing catheters or a power injector is necessary to achieve adequate stent expansion. Analysis of the rupture may identify procedural deficiencies that can be rectified. Conclusions: This technique is simple and timesaving, and interventionists should maintain vigilance when handling of balloon-expandable stents.
Subject
Cardiology and Cardiovascular Medicine,Radiology, Nuclear Medicine and imaging,Surgery
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. Revascularization;Image-Guided Interventions;2020