In Vitro Evaluation of the Gore Viabahn Balloon-Expandable Stent-Graft for Fenestrated Endovascular Aortic Repair

Author:

Torsello Giovanni Federico12ORCID,Herten Monika23,Müller Markus4,Frank André-4,Torsello Giovanni B.2,Austermann Martin2

Affiliation:

1. Department of Diagnostic and Interventional Radiology and Nuclear Medicine, Charité Campus Virchow-Klinikum, Charité University Medicine Berlin, Germany

2. Department of Vascular Surgery, St Franziskus Hospital Münster, Germany

3. Department of Orthopedic and Trauma Surgery, Essen University Hospital, Essen, Germany

4. Biomechanics Laboratory, Department for Hand, Trauma and Reconstructive Surgery, University Hospital Münster, Germany

Abstract

Purpose: To evaluate the safety and integrity of a new stent-graft as a potential bridging device in fenestrated stent-grafts using an in vitro fenestrated model. Materials and Methods: Polyester test sheets with ten 6-mm- or 8-mm-diameter fenestrations were used to simulate a fenestrated main body endoprosthesis. In total, 50 Viabahn balloon-expandable (VBX) stent-grafts of varying lengths (29 and 39 mm) and diameters (6, 7, and 8 mm) were implanted in the fitting fenestrations. After release, the 6- and 7-mm-diameter stent-grafts were flared with a 10×20-mm angioplasty balloon; a 12×20-mm balloon was used in the 8-mm-diameter devices. Safety of the devices was defined as absence of fractures detected on radiography or computed tomography (CT), as well as material failure detected by microscopy and water permeability testing. The forces (in Newtons, N) needed for perpendicular dislocation (pullout force) and axial dislocation (shear stress force) were also evaluated. Results: Forty VBX stent-grafts were subjected to digital radiographic imaging and multiplanar CT. None showed any stent fracture. Subsequent microscopy indicated no damage to the fabric or separation of the graft after flaring. Ten VBX stent-grafts underwent water permeability testing after flaring; no water passed through the graft wall during a 10-minute period under an intraluminal pressure at 120 mm Hg. Testing of 25 VBX stent-grafts revealed initial pullout forces between 11.3 and 31 N. Shear stress tests showed that the average force needed to dislocate the stent-grafts by 50% of their diameter ranged between 5.75 and 6.91 N (mean 6.1±0.5 N) for the 6-mm stents and between 3.31 and 5.4 N (mean 4.4±0.8) for the 8-mm stents. Conclusion: This preliminary study demonstrated the applicability of the VBX as a bridging stent-graft in a simulated fenestration model. A comparison with other stent-grafts and clinical assessment are required.

Funder

W.L. Gore and Associates, Flagstaff, AZ, USA

Cook Medical Inc, Bloomington, IN, USA

Publisher

SAGE Publications

Subject

Cardiology and Cardiovascular Medicine,Radiology, Nuclear Medicine and imaging,Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3