Affiliation:
1. Mercy Medical Center, Pittsburgh, Pennsylvania, USA
Abstract
Purpose: To report a late complication associated with embolization coils used to treat an endoleak after endovascular abdominal aortic aneurysm (AAA) repair. Case Report: A 79-year-old man with a 5.8-cm AAA underwent endovascular repair with an Ancure graft in 1997. A persistent type I endoleak was identified on serial postoperative computed tomographic scans. Three transarterial coil embolization procedures were performed to treat an endoleak from the proximal and right distal attachment sites with outflow by the inferior mesenteric and lumbar arteries. Coil embolization was ultimately successful in sealing the endoleak, and the AAA decreased in size. Four years later, the patient developed an aortoenteric fistula due to erosion of the metallic embolization coils into the duodenum. The endograft was explanted and an extra-anatomical bypass inserted. Conclusions: Coil embolization to treat endoleaks can, on rare occasions, be the cause of aortoenteric fistula. Lifelong follow-up of stent-graft patients is required.
Subject
Cardiology and Cardiovascular Medicine,Radiology, Nuclear Medicine and imaging,Surgery
Cited by
20 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献